Ultrasonic Thickness Gauge NOVOTEST UT-1M-ST

Ultrasonic Thickness Gauge NOVOTEST UT-1M-ST is designed for measuring the thickness of various materials and products with one-sided access. Thickness gauge is also used to measure the thickness of various products from all kinds of plastics, polymers, glasses. NOVOTEST UT-1M-ST is a special version of the ultrasonic thickness gauge NOVOTEST UT-1M, it has metal housing (body), which provide additional safety for the electronic unit, so there are more chances that it is not be damaged in circumstances, where the device with classic plastic case will be broken.

 

The device complies with ASTM E797 Standard Practice for Measuring Thickness by Manual Ultrasonic Pulse-Echo Contact Method.

 

It has Echo-Echo mode. Using special Echo-Echo mode, it allows to measure thickness of metals, through paint coatings (up to 1mm).

  • Unique additional protection of metal housing;
  • Unprecedented autonomy of the device – increased up to 200 hours continuous operation time of the thickness gauge;
  • Special rubber lining on the side surfaces of the casing facilitates the convenient retention of the thickness gauge in the user’s hand; 
  • Device has light weight and dimensions, in comparison with similar protected thickness gauges;
  • Saved data can be transmit to PC;
  • Material selection and automatic setting of ultrasound velocity;
  • Large graphic display of high contrast and backlight;
  • Display of measured thickness in mm and inches;
  • Convenient menu in the device;
  • Wide choice of UT probes with preset settings in the device;
  • Modes of statistical processing of measurements;
  • Ability to restore factory calibrations;
  • Ability to adjust all parameters of the acoustic path;
  • Ability of further adjusting the user gain directly from the measurement mode;
  • Wide range of testing;
  • Mode of audible and visual alarm when leaving the established ranges.
Measuring thicknesses range , mm:

  • Probe 10MHz – P112-10-6 / 2
  • Probe 5MHz – P112-5-10 / 2
  • Probe 2,5MHz – P112-2,5-12 / 2
0.8 … 300 or more

  • range 0.8-30 mm
  • range 1-75 mm
  • range 2-300 mm
Probe dimensions, mm:

  • Probe 10MHz – P112-10-6 / 2
  • Probe 5MHz – P112-5-10 / 2
  • Probe 2,5MHz – P112-2,5-12 / 2
  • D12×15 mm
  • D17×20 mm
  • D20×21 mm
Diameter of probe contact area, mm:

  • Probe 10MHz – P112-10-6 / 2
  • Probe 5MHz – P112-5-10 / 2
  • Probe 2,5MHz – P112-2,5-12 / 2
  • 9 mm
  • 14 mm
  • 16 mm
Size of plate:

  • Probe 10MHz – P112-10-6 / 2
  • Probe 5MHz – P112-5-10 / 2
  • Probe 2,5MHz – P112-2,5-12 / 2
  • 6 mm
  • 10 mm
  • 12 mm
Setting range of the ultrasonic velocity, m / s 1000-9999
Discrete readings on the digital indicator, mm 0.1
Response time, with no more than 1
Basic measurement accuracy, mm ± (0,001 T+/- 0.1)
Echo-Echo mode (through coating measuring), max thickness of coating, mm 1
Standards ASTM E797
Overall dimensions, mm 120x60x25
Operating temperature range, ° C -5 to +40
Power supply 3 AA batteries
Time of continuous work hours, not less, h 10
Weight of electronic unit with battery, no more, kg 0.2
  • Electronic unit Ultrasonic Thickness Gauge UT-1M-ST
  • Transducer (probe) – 1 PC (depends on the desired range of controlled thickness)
  • 3 AA batteries
  • Charger
  • Cables
  • Operating manual
  • Case